{"id":"https://openalex.org/W2004226297","doi":"https://doi.org/10.1088/0026-1394/42/6/s15","title":"Metrological characterization of a new standard, in absolute and gauge pressure modes, in the range 1 Pa to 15 000 Pa","display_name":"Metrological characterization of a new standard, in absolute and gauge pressure modes, in the range 1 Pa to 15 000 Pa","publication_year":2005,"publication_date":"2005-11-28","ids":{"openalex":"https://openalex.org/W2004226297","doi":"https://doi.org/10.1088/0026-1394/42/6/s15","mag":"2004226297"},"language":"en","primary_location":{"id":"doi:10.1088/0026-1394/42/6/s15","doi":"https://doi.org/10.1088/0026-1394/42/6/s15","is_oa":true,"landing_page_url":"https://doi.org/10.1088/0026-1394/42/6/s15","pdf_url":"https://iopscience.iop.org/article/10.1088/0026-1394/42/6/S15/pdf","source":{"id":"https://openalex.org/S7291761","display_name":"Metrologia","issn_l":"0026-1394","issn":["0026-1394","1681-7575"],"is_oa":false,"is_in_doaj":false,"is_core":true,"listed_in":["anvur-scientific","cwts-core","dongbi-b","jufo-2","norway-1","poland-70","russia-white-list-2"],"host_organization":"https://openalex.org/P4310320083","host_organization_name":"IOP Publishing","host_organization_lineage":["https://openalex.org/P4310320083","https://openalex.org/P4310311669"],"host_organization_lineage_names":["IOP Publishing","Institute of Physics"],"type":"journal"},"license":null,"license_id":null,"version":"publishedVersion","is_accepted":true,"is_published":true,"raw_source_name":"Metrologia","raw_type":"journal-article"},"volume":null,"type":"article","is_traditional":true,"indexed_in":["crossref"],"open_access":{"is_oa":true,"oa_status":"bronze","oa_url":"https://iopscience.iop.org/article/10.1088/0026-1394/42/6/S15/pdf","any_repository_has_fulltext":false},"authorships":[{"author_position":"first","author":{"id":"https://openalex.org/A5003682544","display_name":"Pierre Otal","orcid":"https://orcid.org/0000-0002-5415-8984","observed_orcids":["https://orcid.org/0000-0002-5415-8984"]},"institutions":[{"id":"https://openalex.org/I4210107682","display_name":"Laboratoire National de M\u00e9trologie et d'Essais","ror":"https://ror.org/01ph39d13","country_code":"FR","type":"other","lineage":["https://openalex.org/I4210107682"]}],"countries":["FR"],"is_corresponding":false,"raw_author_name":"P Otal","raw_affiliation_strings":["LNE, 1, rue Gaston Boissier, 75724 Paris Cedex 15, France"],"raw_orcid":null,"affiliations":[{"raw_affiliation_string":"LNE, 1, rue Gaston Boissier, 75724 Paris Cedex 15, France","institution_ids":["https://openalex.org/I4210107682"]}]},{"author_position":"last","author":{"id":"https://openalex.org/A5034557718","display_name":"J C Legras","orcid":null,"observed_orcids":[]},"institutions":[{"id":"https://openalex.org/I4210107682","display_name":"Laboratoire National de M\u00e9trologie et d'Essais","ror":"https://ror.org/01ph39d13","country_code":"FR","type":"other","lineage":["https://openalex.org/I4210107682"]}],"countries":["FR"],"is_corresponding":false,"raw_author_name":"J C Legras","raw_affiliation_strings":["LNE, 1, rue Gaston Boissier, 75724 Paris Cedex 15, France"],"raw_orcid":null,"affiliations":[{"raw_affiliation_string":"LNE, 1, rue Gaston Boissier, 75724 Paris Cedex 15, France","institution_ids":["https://openalex.org/I4210107682"]}]}],"institutions":[],"countries_distinct_count":1,"institutions_distinct_count":1,"corresponding_author_ids":[],"corresponding_institution_ids":["https://openalex.org/I4210107682"],"apc_list":{"value":3325,"currency":"USD","value_usd":3325},"apc_paid":null,"fwci":0.0,"has_fulltext":true,"cited_by_count":6,"citation_normalized_percentile":{"value":0.0,"is_in_top_1_percent":false,"is_in_top_10_percent":false},"cited_by_percentile_year":{"min":87,"max":96},"biblio":{"volume":"42","issue":"6","first_page":"S216","last_page":"S219"},"is_retracted":false,"is_paratext":false,"is_xpac":false,"primary_topic":{"id":"https://openalex.org/T11890","display_name":"Scientific Measurement and Uncertainty Evaluation","score":0.9994999766349792,"subfield":{"id":"https://openalex.org/subfields/3105","display_name":"Instrumentation"},"field":{"id":"https://openalex.org/fields/31","display_name":"Physics and Astronomy"},"domain":{"id":"https://openalex.org/domains/3","display_name":"Physical Sciences"}},"topics":[{"id":"https://openalex.org/T11890","display_name":"Scientific Measurement and Uncertainty Evaluation","score":0.9994999766349792,"subfield":{"id":"https://openalex.org/subfields/3105","display_name":"Instrumentation"},"field":{"id":"https://openalex.org/fields/31","display_name":"Physics and Astronomy"},"domain":{"id":"https://openalex.org/domains/3","display_name":"Physical Sciences"}},{"id":"https://openalex.org/T11583","display_name":"Advanced Measurement and Metrology Techniques","score":0.00019999999494757503,"subfield":{"id":"https://openalex.org/subfields/2210","display_name":"Mechanical Engineering"},"field":{"id":"https://openalex.org/fields/22","display_name":"Engineering"},"domain":{"id":"https://openalex.org/domains/3","display_name":"Physical Sciences"}},{"id":"https://openalex.org/T12019","display_name":"Calibration and Measurement Techniques","score":0.00019999999494757503,"subfield":{"id":"https://openalex.org/subfields/2202","display_name":"Aerospace Engineering"},"field":{"id":"https://openalex.org/fields/22","display_name":"Engineering"},"domain":{"id":"https://openalex.org/domains/3","display_name":"Physical Sciences"}}],"keywords":[{"id":"https://openalex.org/keywords/pressure-standards","display_name":"pressure standards","score":0.6570000052452087},{"id":"https://openalex.org/keywords/capacitance-diaphragm-gauge","display_name":"capacitance diaphragm gauge","score":0.7139999866485596},{"id":"https://openalex.org/keywords/gauge-pressure","display_name":"gauge pressure","score":0.6980000138282776},{"id":"https://openalex.org/keywords/absolute-pressure","display_name":"absolute pressure","score":0.996999979019165},{"id":"https://openalex.org/keywords/effective-area","display_name":"effective area","score":0.8360000252723694},{"id":"https://openalex.org/keywords/correction-factors","display_name":"correction factors","score":0.6980000138282776},{"id":"https://openalex.org/keywords/measurement-uncertainty","display_name":"measurement uncertainty","score":0.609000027179718},{"id":"https://openalex.org/keywords/repeatability","display_name":"repeatability","score":0.875},{"id":"https://openalex.org/keywords/linearity","display_name":"linearity","score":0.9810000061988831},{"id":"https://openalex.org/keywords/low-pressure-measurement","display_name":"low-pressure measurement","score":0.4169999957084656}],"concepts":[{"id":"https://openalex.org/C195766429","wikidata":"https://www.wikidata.org/wiki/Q394","display_name":"Metrology","level":2,"score":0.7365588545799255},{"id":"https://openalex.org/C40976572","wikidata":"https://www.wikidata.org/wiki/Q2330873","display_name":"Gauge (firearms)","level":2,"score":0.6884082555770874},{"id":"https://openalex.org/C80264047","wikidata":"https://www.wikidata.org/wiki/Q7424019","display_name":"Pressure measurement","level":2,"score":0.5917367935180664},{"id":"https://openalex.org/C204323151","wikidata":"https://www.wikidata.org/wiki/Q905424","display_name":"Range (aeronautics)","level":2,"score":0.5187511444091797},{"id":"https://openalex.org/C2780841128","wikidata":"https://www.wikidata.org/wiki/Q5073781","display_name":"Characterization (materials science)","level":2,"score":0.5058133602142334},{"id":"https://openalex.org/C2777487392","wikidata":"https://www.wikidata.org/wiki/Q481201","display_name":"Absolute (philosophy)","level":2,"score":0.4632740020751953},{"id":"https://openalex.org/C121332964","wikidata":"https://www.wikidata.org/wiki/Q413","display_name":"Physics","level":0,"score":0.40182992815971375},{"id":"https://openalex.org/C192562407","wikidata":"https://www.wikidata.org/wiki/Q228736","display_name":"Materials science","level":0,"score":0.3671610355377197},{"id":"https://openalex.org/C97355855","wikidata":"https://www.wikidata.org/wiki/Q11473","display_name":"Thermodynamics","level":1,"score":0.2286105751991272},{"id":"https://openalex.org/C120665830","wikidata":"https://www.wikidata.org/wiki/Q14620","display_name":"Optics","level":1,"score":0.15795010328292847},{"id":"https://openalex.org/C138885662","wikidata":"https://www.wikidata.org/wiki/Q5891","display_name":"Philosophy","level":0,"score":0.10105019807815552},{"id":"https://openalex.org/C159985019","wikidata":"https://www.wikidata.org/wiki/Q181790","display_name":"Composite material","level":1,"score":0.07660093903541565},{"id":"https://openalex.org/C191897082","wikidata":"https://www.wikidata.org/wiki/Q11467","display_name":"Metallurgy","level":1,"score":0.06144976615905762},{"id":"https://openalex.org/C111472728","wikidata":"https://www.wikidata.org/wiki/Q9471","display_name":"Epistemology","level":1,"score":0.05897176265716553}],"mesh":[],"locations_count":1,"locations":[{"id":"doi:10.1088/0026-1394/42/6/s15","doi":"https://doi.org/10.1088/0026-1394/42/6/s15","is_oa":true,"landing_page_url":"https://doi.org/10.1088/0026-1394/42/6/s15","pdf_url":"https://iopscience.iop.org/article/10.1088/0026-1394/42/6/S15/pdf","source":{"id":"https://openalex.org/S7291761","display_name":"Metrologia","issn_l":"0026-1394","issn":["0026-1394","1681-7575"],"is_oa":false,"is_in_doaj":false,"is_core":true,"listed_in":["anvur-scientific","cwts-core","dongbi-b","jufo-2","norway-1","poland-70","russia-white-list-2"],"host_organization":"https://openalex.org/P4310320083","host_organization_name":"IOP Publishing","host_organization_lineage":["https://openalex.org/P4310320083","https://openalex.org/P4310311669"],"host_organization_lineage_names":["IOP Publishing","Institute of Physics"],"type":"journal"},"license":null,"license_id":null,"version":"publishedVersion","is_accepted":true,"is_published":true,"raw_source_name":"Metrologia","raw_type":"journal-article"}],"best_oa_location":{"id":"doi:10.1088/0026-1394/42/6/s15","doi":"https://doi.org/10.1088/0026-1394/42/6/s15","is_oa":true,"landing_page_url":"https://doi.org/10.1088/0026-1394/42/6/s15","pdf_url":"https://iopscience.iop.org/article/10.1088/0026-1394/42/6/S15/pdf","source":{"id":"https://openalex.org/S7291761","display_name":"Metrologia","issn_l":"0026-1394","issn":["0026-1394","1681-7575"],"is_oa":false,"is_in_doaj":false,"is_core":true,"listed_in":["anvur-scientific","cwts-core","dongbi-b","jufo-2","norway-1","poland-70","russia-white-list-2"],"host_organization":"https://openalex.org/P4310320083","host_organization_name":"IOP Publishing","host_organization_lineage":["https://openalex.org/P4310320083","https://openalex.org/P4310311669"],"host_organization_lineage_names":["IOP Publishing","Institute of Physics"],"type":"journal"},"license":null,"license_id":null,"version":"publishedVersion","is_accepted":true,"is_published":true,"raw_source_name":"Metrologia","raw_type":"journal-article"},"sustainable_development_goals":[],"study_designs":[],"awards":[],"funders":[],"has_content":{"pdf":true,"grobid_xml":true},"content_urls":{"pdf":"https://content.openalex.org/works/W2004226297.pdf","grobid_xml":"https://content.openalex.org/works/W2004226297.grobid-xml"},"referenced_works_count":1,"referenced_works":["https://openalex.org/W2086862492"],"related_works":["https://openalex.org/W2765900014","https://openalex.org/W2795036553","https://openalex.org/W2034966011","https://openalex.org/W1975210168","https://openalex.org/W2057819326","https://openalex.org/W2333777007","https://openalex.org/W2118101271","https://openalex.org/W1990956588","https://openalex.org/W2092761917","https://openalex.org/W2071907761"],"abstract_inverted_index":{"The":[0,47,93,107,129],"metrological":[1],"characterization":[2],"of":[3,100,124],"a":[4,39],"new":[5],"standard":[6,114],"in":[7,13,30,50,142,154],"absolute":[8,33,155],"and":[9,32,102,145],"gauge":[10,31,143],"pressure":[11,45,108],"modes,":[12],"the":[14,51,55,71,89,122],"range":[15],"1":[16],"Pa":[17],"to":[18,77,85],"15":[19],"000":[20],"Pa,":[21],"is":[22,58,115,133],"described.":[23],"Its":[24],"effective":[25,52],"area":[26,53],"has":[27,67,95],"been":[28,68,83],"determined":[29],"modes":[34,57,73],"by":[35],"direct":[36],"comparison":[37],"with":[38,118],"20":[40],"cm":[41],"2":[42],"simple":[43],"piston":[44],"balance.":[46],"relative":[48],"difference":[49],"between":[54,70],"two":[56,72],"1.2":[59],"\u00d7":[60,137,140,149,152],"10":[61,138,150],"\u22126":[62],".":[63],"No":[64],"significant":[65],"effect":[66],"observed":[69],"for":[74,112],"pressures":[75],"down":[76],"5":[78],"kPa.":[79],"Correction":[80],"coefficients":[81],"have":[82],"calculated":[84],"take":[86],"into":[87],"account":[88],"initial":[90],"condition":[91],"changes.":[92],"evaluation":[94],"also":[96],"provided":[97],"an":[98],"estimate":[99],"repeatability":[101],"linearity":[103],"at":[104],"low":[105],"pressure.":[106],"measurement":[109],"uncertainty":[110,132],"budget":[111],"this":[113],"presented":[116],"along":[117],"some":[119],"results":[120],"from":[121],"calibration":[123],"working":[125],"capacitance":[126],"diaphragm":[127],"gauges.":[128],"estimated":[130],"expanded":[131],"8":[134],"mPa+":[135,147],"2.4":[136],"\u22125":[139,151],"p":[141,153],"mode":[144],"13":[146],"2.3":[148],"mode.":[156]},"counts_by_year":[{"year":2022,"cited_by_count":1},{"year":2021,"cited_by_count":1},{"year":2020,"cited_by_count":1},{"year":2018,"cited_by_count":2},{"year":2016,"cited_by_count":1}],"updated_date":"2026-11-07T11:25:07.637180","created_date":"2016-06-24T00:00:00"}